Floor-mounted distribution cabinets
Application
The IT system distribution cabinets of the VIT-C series supply electrical power to group 2 medical areas, e.g. operating theatres. In these areas the protective measure "Indication by insulation monitoring in the IT system" must be used for the socket-outlet circuits for electrical medical equipment, circuits for operating theatre luminaires and similar lighting supplied with nominal voltages exceeding AC 25 V or DC 60 V. In addition, a changeover module is required to change over automatically to a second independent supply of power, SV (safety power supply source)/AV (normal power supply source) or UPS/SV (safety power supply source), in case of failure of the primary cable.
Device features
Internal components
• Isolating transformer (3150 to 8000 VA)
• Changeover module with voltage monitoring
• IT system monitoring
• 6 circuits with 2-pole circuit breakers B16A
• Customers can save time and money as the distribution cabinets are
supplied prewired
• Covering of sheet-steel according to
DIN VDE 0100-710:2002-11 para. 710.51.2.2
• Designed in accordance with the requirements of applicable standards
• Voluntary test of the changeover module by TÜV (German Technical
Inspection Authority)
• Connection for 3 annunciator combinations MK2430
• Connection for TM annunciator and control panels
• Variable changeover period t = 0.5 ... 20 s
• Screwless-type terminals
• Exchange of information via bus technology
• Short delivery times
Individual, project-specific requirements can be taken into account at any time.
Wall-mounted distribution cabinets
Application
The IT system wall-mounted distribution cabinets supply electrical power to group 2 medical areas, e.g. operating theatres. In these areas the protective measure "Indication by insulation monitoring in the IT system" must be used for the socket-outlet circuits for electrical medical equipment, circuits for operating theatre luminaires and similar lighting supplied with nominal voltages exceeding AC 25 V or DC 60 V. In addition, a changeover module is required to change over automatically to a second independent supply of power, SV (safety power supply source)/AV (normal power supply source) or UPS/SV (safety power supply source), in case of failure of the primary cable.
Device features
Internal components
• Changeover module with voltage monitoring
• IT system monitoring
• 6 circuits with 2-pole circuit breakers B16A
• Customers can save time and money as the distribution cabinets are
supplied prewired
• Covering of sheet-steel according to DIN VDE 0100-710:2002-11
para. 710.51.2.2
• Designed in accordance with the requirements of applicable standards
• The isolating transformer (3150 to 8000 VA) is housed externally to the
wall-mounted distribution cabinet.
• Voluntary test of the changeover module by TÜV (German Technical
Inspection Authority)
• Connection for 3 annunciator combinations MK2430
• Connection for TM annunciator and control panels
• Variable switchover period t = 0.5 ... 20 s
• Screwless-type terminals
• Exchange of information via bus technology
• Short delivery times
Individual, project-specific requirements can be taken into account at any time.